I
have already written extensively in this blog about Mass Spectrometry
here
So
what I want to do in this blog is take mass spectrometry (MS) to the next level
as it were.
I
am going to discuss the “time of flight” mass spectrometer.
I
am also going to discuss how mass spectrometry is used to analyse organic molecules
– the organic molecules separated out from a sample using gas chromatography
(GC).
I
am also going to discuss how gas chromatography works.
Time of Flight Mass
Spectrometer
This
is the simplest type of mass spectrometer and they have been sent on space
missions such as that on the Philae lander that made the first soft landing on
a moving comet in November 2014.
How
does this type of mass spectrometer work?
The
sample is injected into the machine and it is subjected to fast electron
bombardment.
An
electric field then accelerates (it actually pushes the positive ions since the
ions experience a positive field) the resultant positive ions towards a negatively charged plate but the ions do
not then enter a magnetic field instead they enter a long tube or flight tube
(see diagram) which is about 0.6—2.0m in length.
All
ions acquire the same kinetic energy but some travel faster because they are
lighter than others.
This
tube is field free and it is the momentum of the ions that means they reach the
detector.
So
the lighter ions have greater velocity than the heavier ions.
The
lighter ions take a shorter time to travel down the tube so that they are
detected first at the ion detector.
Other
heavier ions take a longer time to travel to the detector.
Flight
times are proportional to the square root of the mass of the positive ion
Flight
time = drift length = drift length × √(mass/2×energy)
velocity
Knowing
the flight time means the mass of that positive particle can be determined and
a mass spectrum plotted.
As
a result, the different fragment ions from the sample can be determined.
For
example, a mass spectrometer with a 0.6m flight tube will mean that a molecule
of mass 26amu will take 6 × 10–6 seconds to complete the flight
path.
For
good distinction between different masses, the instrument’s timing electronics
must be capable of measuring the time of flight to a resolution of 0.25 ns
(0.25 x 10-9 seconds).
Ions
of different mass arrive at the detector one after the other so that it is
possible to detect all the ions that were derived from the source.
Ions are detected by gaining electrons.
The abundance of any one particular isotope is measured by the size of the current flowing in the detector.
Ions are detected by gaining electrons.
The abundance of any one particular isotope is measured by the size of the current flowing in the detector.
This
is why a time of flight mass spectrometer can have a very high overall
sensitivity compared to other MS systems.
It
also means that time of flight spectrometers have no difficulties with high
mass ions; it is just a matter of waiting for them to arrive.
The
high mass limit is then determined by the ability of the detector to register the
arrival of the high mass ion.
This
means they can determine the mass of large molecules such as proteins with some
degree of accuracy say 0.01%
If you are studying AQA A level there are extensive notes on Time of Flight mass spectrometers here .
If you are studying AQA A level there are extensive notes on Time of Flight mass spectrometers here .
The
time of flight instrument on the Philae lander was coupled to a Gas Chromatograph.
This
is a common analytical combination.
Here
is how the gas Chromatograph operates
Gas Chromatograph
It
works like this:
The
chromatograph column contains an inert solid and it is between 1.5 and 10m long
with an internal diameter of about 4mm.
On
the surface of the powder in this column there is adsorbed a liquid called the
liquid stationary phase because it does not move.
A
chemically inert gas is flowing through the column–gases used include carbon
dioxide, nitrogen and argon–the choice of gas depends on the sample under test
and the kind of detector being used.
The
sample under test is vaporised and injected into the column.
As
the sample mixture passes along the column, some parts of it are retained on
the column for longer than others.
The
separated components of the mixture emerge from the column according to their
retention time (Rt )
It’s
very important to prevent molecules like water entering the column.
A
small sample is usually used, say 20 μlitres, and injected as a “plug” of vapour.
There is usually a flash vaporizer at the head of the
column that operates at about 50oC above the known boiling point of
the least volatile component in the sample.
You
will see in the diagram that the column is in an oven and the temperature of
the oven is controlled very precisely to within tenths of a degree
Celsius.
The
time taken for the sample to pass through the column is called the elution
time.
If
the GC is linked to a Mass Spectrometer as on the Philae lander then the sample
is directly transferred to the MS after separation in the GC.
Other
wise a flame, ionisation detector picks up the different samples from the
mixture.
The
output from the column is mixed with hydrogen and air, and ignited.
Organic
compounds burning in the flame produce ions and electrons.
The
electrons can conduct electricity through the flame.
The
resultant current is measured.
The
only problem with this approach is that the sample is destroyed.
If
the samples are put through a mass spectrometer they are also destroyed.
How to analyse the mass
spectra of organic molecules.
So
you have analysed your organic sample in the Gas Chromatograph and each of the
components of your sample mixture has been put through the Mass Spectrometer.
Your
Gas Chromatogram might look like this::
It
looks like a drugs test sample: note the amphetamine peak.
Note
the peak at 2.055ms.
The
individual components are identified by their retention times compared with
standards already known.
Note
the horizontal scale is a labeled “time” i.e. retention time on the
column.
And
the Mass Spectrum of one of the samples in the mixture might look like this:
But
how do you analyse mass spectra like this?
The
mass spectrum reveals the fragmentation pattern of the molecule on bombardment
with high energy electrons.
Take
this sample of an alkane:
You
can find most molecules at the site above: http://webbook.nist.gov/chemistry
First,
note the highest mass peak at 142.
This
is the M+ peak or molecular
ion or parent ion peak.
It
represents the ionised decane molecule or the
C10H22+ ion.
Now
I want you to see the pattern that follows in the masses of the higher peaks
that occur in little clusters.
These
masses are:
113
99
85
71
57
43
First
note how each is separated by 14 mass units: that's a —CH2— or
methylene group.
You
can see how this molecules a linear alkane is fragmenting as —CH2—
groups are broken off one at a time.
So
what positive ion does each mass represent?
113
is 142–29 so it must be C8H17+
99
is 113–14 so it must be C7H15+
85
is 99-14 so it must be C6H13+
71
is 85–14 so it must be C5H11+
57
is 71– 14 so it must be C4H9+
and
43 is 57-14 so it must be C3H7+ etc…….
This
is the distinctive pattern of fragmentation of a linear alkane molecule or
alkane type side chain in a molecule.
Lastly,
note too the very small peak at m/z 143
It
is a tenth the size of the parent ion peak and represents the abundance of the
molecular ion with one C13 carbon isotope in it.
e.g.
the C9C13H22 +
ion.
Lets
look now at a typical halogenoalkane: chloroethane C2H5Cl
Again
we can see the alkane fragmentation pattern but the significant thing here is
the signature of chlorine.
Look
at the peaks at m/z 66 and 64
66
is one third the height of 64
This
reflects the abundance of chlorine isotopes in the environment where Cl35
is three times more abundant than Cl37
Both
these peaks represent molecular or parent ions
66
is C2H5Cl37
and 64 is C2H5Cl35
Here
is a neat way if detecting the presence of one chlorine atom in a molecule.
This
is a likely question in a college or school advanced exam paper.
Knowing
how alkanes fragment you should be able to work out what ions the other major
peaks represent.
Here
is another mass spectrum this time it's a bromo compound
This
is the mass spectrum of bromoethane.
What
is distinctive about this mass spectrum is that there are two parent or
molecular ion peaks at m/z 108 and 110.
These
two peaks reflect the relative abundance in nature of the two bromine isotopes
Br79 and Br81.
You can see the two very small peaks at 79 and 81 showing the bromine atoms that have fragmented as cations from the molecule.
The
isotopes are about equally abundant so the two parent ion peaks are of about
the same intensity.
Here
is another signature mark on a mass spectrum that tells you there is one
bromine atom in the molecule.
Here
is another mass spectrum this time of an aromatic compound
This
is the mass spectrum of benzyl alcohol: C6H5CH2OH
The
characteristic feature of the mass spectra of aromatics is that they usually
carry a peak at 77 as you can see here.
Why
77?
77
corresponds to a phenyl ion C6H5+
Aromatic
molecules tend to fragment by the ring substituents breaking off one at a
time.
This
leaves the phenyl ring at 77.
updated 30th October 2017
updated 30th October 2017